When using a standard 5 lead electrode placement, the electrode for the la should be located at:

3 lead ecg placement and preparation

Skin preparation

Proper skin prep is essential to obtain good quality tracing:

  • Choose designated chest electrode placement sites
  • Avoid placing electrodes over bony prominences i.e. clavicles
  • To enhance electrode skin contact and patient comfort:
    • Clip /shave excess chest hair in a 1-2 cm diameter of the designated electrode site, if necessary, with the patients consent
    • With a dry swab, gently clean /abrade electrode sites to remove loose skin cells, body oil, sweat and improve skin/electrode contact
    • If alcohol pad is used to cleanse skin, allow to site to dry, before electrode placement. ECG Electrodes are small sensor pads (self adhesive,disposable, pre gelled) applied on the skin to enable detection of the electrical activity of the heart, which is transmitted to the monitor, amplified and displayed as the ECG trace

When using a standard 5 lead electrode placement, the electrode for the la should be located at:

3 lead ECG cable Placement (there are two ways)

Way 1

Monitors one of the three leads:

  • RA: placed the red electrode within the frame of rib cage,right under the clavicle near  shoulder( see chart in follow picture)
  • LA: the yellow electrode is placed below  left clavicle, which is in the same level of the Red electrode
  • LL: green electrode is located on the left side,under the  pectoral muscles, lower edge of left rib cage

When using a standard 5 lead electrode placement, the electrode for the la should be located at:

Way 2

Arrange the placement of the 3 electrode according to Eindhoven,s Triangle:

RA:place the red electrode at the Atrial Depolarization

LA:place the LA electrode at the right arm,under the right shoulder Clavicle

LL:Left leg

Note:different manufacturer may define didfferent elctrodes in different color.Please follow the manucturer,s instructure to place.

When using a standard 5 lead electrode placement, the electrode for the la should be located at:

3 lead ecg placement tutor video

technical specification

Items Specification
Products name  
Model EC203S
Conductor  Material:Tinned or Silvered Tinsel+KEVLAR
 Specification:7/0.23+200D KEVLAR
 Insulation  Material:HDPE
 Nom.Thickness:0.34
 Nom.O.D.:ø01.40士0.05
 Colour:White/Black/Red
 Conducting Layer  Material:Semiconductive PVC
 Nom.Thickness:0.1
 Specificationø1.6士0.08
Colour:Black
 Shield  Material:Tinned copper
 Specification:50士3/0.10TC
 Coverage:>90%
 Jacket  MaterialTPU
 Nom.Thickness:0.72
 Nom.O.D.:ø3.2士0.10/9.6士0.20
 Colour:Matte Gray

Where should LL and RL electrodes be placed?

Place LL (red) electrode on the lower left abdomen within the rib cage frame. Place RL (green) electrode on lower right abdomen within the rib cage frame.

Where is V5 lead placed?

To place the electrode for lead V5 start in the intercostal space associated with lead V4 (5th intercostal space) and move to the left to an imaginary line associated with the front portion of the armpit going down toward the anterior hip. Hence, V5 is in the 5th intercostal space, anterior axillary line.