Which of the following instructions should a medical assistant provide to a patient prior to spirometry testing?

Which of the following instructions should a medical assistant provide to a patient prior to spirometry testing?

, R. Crapo, J. Hankinson, V. Brusasco, F. Burgos, R. Casaburi, A. Coates, P. Enright, C. P. M van der Grinten, P. Gustafsson, R. Jensen, D. C. Johnson, N. MacIntyre, R. McKay, D. Navajas, O. F. Pedersen, R. Pellegrino, G. Viegi, J. Wanger

European Respiratory Journal 2005 26: 153-161; DOI: 10.1183/09031936.05.00034505

  • Diffusing capacity
  • infections
  • lung function measurements
  • lung volume measurements
  • reference values
  • spirometry standardisation

BACKGROUND

In preparing the joint statements on lung function testing for the American Thoracic Society (ATS) and the European Respiratory Society (ERS), it was agreed by the working party that the format of the statements should be modified so that they were easier to use by both technical and clinical staff. This statement contains details about procedures that are common for many methods of lung function testing and, hence, are presented on their own. A list of abbreviations used in all the documents is also included as part of this statement.

DEFINITIONS

All terms and abbreviations used here are based on a report of the American College of Chest Physicians/ATS Joint Committee on Pulmonary Nomenclature 1. The metrology definitions agreed by the International Standards Organization (ISO) are recommended 2 and some important terms are defined as follows.

Accuracy is the closeness of agreement between the result of a measurement and the conventional true value.

Repeatability is the closeness of agreement between the results of successive measurements of the same item carried out, subject to all of the following conditions: same method, same observer, same instrument, same location, same condition of use, and repeated over a short space of time. In previous documents, the term reproducibility was used in this context, and this represents a change towards bringing this document in line with the ISO.

Reproducibility is the closeness of agreement of the results of successive measurements of the same item where the individual measurements are carried out with changed conditions, such as: method of measurement, observer, instrument, location, conditions of use, and time. Thus, if a technician tests a subject several times, this is looking at the repeatability of the test. If the subject is then given a bronchodilator drug and tested again after 30 min, one needs to know the reproducibility of the test in order to make a decision on this comparison.

The measurement range for a recording device is the range over which the manufacturer indicates the device complies with the recommendations below.

Equipment resolution is the smallest detectable change in measurement.

PATIENT CONSIDERATIONS

Contraindications

Performing lung function tests can be physically demanding for a minority of patients. It is recommended that patients should not be tested within 1 month of a myocardial infarction. Patients with any of the conditions listed in table 1 are unlikely to achieve optimal or repeatable results.

Position

Testing may be performed either in the sitting or standing position, and the position should be recorded on the report 3, 4. Sitting is preferable for safety reasons in order to avoid falling due to syncope. The chair should have arms and be without wheels. If a wheelchair is used, the wheels should be locked. If the standing position is used, a chair can be placed behind the patient/subject, so that they can be quickly and easily moved into a sitting position if they become light-headed during the manoeuvre. Obese subjects, or those with excessive weight at the mid-section, will frequently obtain a deeper inspiration when tested in the standing position. Consequently, forced expiratory volumes and flows may improve with the standing position in these individuals. Normal-weight subjects typically have equivalent values when tested sitting or standing, but, for longitudinal studies, the same test position should be used each time.

PATIENT DETAILS

Age, height and weight

The patient's age, height and weight (wearing indoor clothes without shoes) are recorded for use in the calculation of reference values. The age should be expressed in years. Height and weight should be expressed with the units in use in the country, corresponding to the ones of the selected reference equation. Body mass index should be calculated as kg·m−2. The height should be measured without shoes, with the feet together, standing as tall as possible with the eyes level and looking straight ahead, and using an accurate measuring device. For patients with a deformity of the thoracic cage, such as kyphoscoliosis, the arm span from fingertip to fingertip can be used as an estimate of height. Arm span should be measured with the subject standing against a wall with the arms stretched to attain the maximal distance between the tips of the middle fingers. A regression equation using arm span, race, sex and age has been found to account for 87% of the variance in standing height 5, with the standard error of the estimate for height ranging from 3.0 to 3.7 cm. Using fixed arm-span ratios (e.g. height = arm span/1.06) estimated the standing height reasonably well, except at the extremes, but was always inferior to the regression equation. Estimating height in this way introduces a further level of uncertainty with regard to the predicted value of the lung function index, and the use of fixed ratios has been shown to lead to misclassification of disease 6. The use of knee height to predict height can also be used for handicapped people where arm span may be difficult to measure 7, 8.

Therapy

The operator should record the type and dosage of any (inhaled or oral) medication that may alter lung function and when the drugs were last administered.

Subject preparation

Subjects should avoid the activities listed in table 2, and these requirements should be given to the patient at the time of making the appointment. On arrival, all of these points should be checked, and any deviations from them recorded.

Subjects should be as relaxed as possible before and during the tests. The decision to avoid long- and short-acting bronchodilators is a clinical one, dependent on the question being asked. If the study is performed to diagnose an underlying lung condition, then avoiding bronchodilators is useful. If the study is carried out to determine a response to an existing therapeutic regimen, then one may choose not to withhold bronchodilator medications.

Patients should be asked to loosen tight-fitting clothing. Dentures should normally be left in place; if they are loose, they may interfere with performance and are, therefore, best removed.

LABORATORY DETAILS

Ambient temperature, barometric pressure and time of day must be recorded. Temperature is an important variable in most pulmonary function tests and is often measured directly by the instrument. The way in which it is measured and used may vary from instrument to instrument. For example, it may be measured with a simple thermometer or an internal thermistor. Regardless of the method used, it is the responsibility of the laboratory to confirm the accuracy of temperature measurements, and it is the responsibility of the manufacturer to describe or provide a clear mechanism for checking the accuracy of instrument temperature measurements. They should also provide instructions on how to respond when acceptable temperature performance cannot be confirmed.

Ideally, when patients return for repeat testing (e.g. at a clinic), the equipment and the operator should be the same, and the time of day should be within 2 h of previous test times.

The order for performing lung function tests should take into account the optimum work flow in the laboratory, potential influences of one test on another and the ability of the subject to undertake the test. One possible order is shown in table 3.

There should be appropriate delays between tests, as indicated in the subsequent sections of this series of documents. Other orders of testing are acceptable (e.g. static lung volumes, diffusing capacity, dynamic studies, inhalation of bronchodilator agent and then repeat dynamic studies, as taken from table 3), but the order should be kept constant to avoid introducing unanticipated variability to test results. The choice of order of testing should consider the potential effect of one test on the subsequent test. For example, the measurement of carbon monoxide diffusing capacity of the lung (DL,CO) immediately after a nitrogen washout measurement of the total lung capacity (TLC) will be affected by the increased oxygen content in the lungs, unless enough time has passed to allow the oxygen concentration to return to normal. Also, tidal breathing manoeuvres may be disturbed by a recently performed maximal forced expiratory manoeuvre. Bronchodilator administration may affect static lung volumes, reducing hyperinflation by up to 0.5 L 9. While bronchodilators do not seem to affect diffusing capacity when measured by the Jones–Meade method, they may allow ∼10% of patients to obtain a measurement of diffusing capacity that was not possible pre-bronchodilator 10.

HYGIENE AND INFECTION CONTROL

The goal of infection control is to prevent the transmission of infection to patients/subjects and staff during pulmonary function testing. The number of documented cases of infection transmission is very small, but the potential is real (see Level of infection risk section). This set of recommendations focuses on equipment used to measure spirometry, diffusing capacity and lung volumes. Organisms may also be transmitted via pulse oximeter probes and nebulisers used to administer bronchodilators 11, 12. Although infection risks increase with exposure to blood, this document does not deal with the risks of arterial blood gases. Pulmonary laboratories performing blood gas analysis should follow the same infection-control procedures used by their clinical laboratory.

Infection can be transmitted by direct contact or by indirect means, which is discussed as follows.

Transmission by direct contact

There is potential for transmission of upper respiratory diseases, enteric infections and blood-borne infections through direct contact. Although hepatitis and HIV contagion are unlikely via saliva, transmission becomes a possibility with open sores on the oral mucosa or bleeding gums. The most likely surfaces for contact are mouthpieces and the immediate proximal surfaces of valves or tubing.

Transmission by indirect contact

There is potential for transmission of tuberculosis (TB), various viral infections, opportunistic infections and nosocomial pneumonia through aerosol droplets. The most likely surfaces for possible contamination by this route are mouthpieces, proximal valves and tubing.

Prevention

Transmission to technicians

Prevention of infection transmission to technicians exposed to contaminated spirometer surfaces can be accomplished through proper hand washing and use of barrier devices, such as suitable gloves. To avoid technician exposure and cross-contamination, hands should be washed immediately after direct handling of mouthpieces, tubing, breathing valves or interior spirometer surfaces. Gloves should be worn when handling potentially contaminated equipment if the technician has any open cuts or sores on his/her hands. Hands should always be washed between patients. Indications and techniques for hand washing during pulmonary function testing have previously been reviewed 13.

Cross-contamination

To avoid cross-contamination, reusable mouthpieces, breathing tubes, valves and manifolds should be disinfected or sterilised regularly. Mouthpieces, nose clips and any other equipment that comes into direct contact with mucosal surfaces should be disinfected, sterilised or, if disposable, discarded after each use. The optimal frequency for disinfection or sterilisation of tubing, valves or manifolds has not been established. However, any equipment surface showing visible condensation from expired air should be disinfected or sterilised before reuse.

Since the use of cold sterilising agents is not without risk, laboratory staff should take care to follow the manufacturer's recommendations concerning proper handling of these products. Some respiratory equipment may be damaged by some methods of sterilisation. For example, heat sterilisation or cold sterilisation chemicals could damage some flow sensors, tubing or seals. Manufacturers should explicitly describe acceptable methods of cleaning and disinfecting their equipment, including recommended chemicals and concentrations, as well as safety precautions for the technicians. Manufacturers' recommendations should be followed; however, a hospital infection-control department's requirements will probably supersede both manufacturers' recommendations and those in this document. If hospital infection-control recommendations have the potential to harm instruments, compromises may have to be negotiated.

Volume-based spirometers

Volume-based spirometers used with a closed-circuit technique should be flushed between subjects with room air at least five times over the entire volume range of the spirometer to enhance clearance of droplet nuclei. The breathing tube and mouthpiece should be decontaminated or changed between patients.

When the open-circuit technique is used and the patient/subject only exhales into the spirometer, only the portion of the circuit through which rebreathing occurs must be decontaminated between patients. For example, when a pneumotachometer system is used, either avoid having the patient inspire through the device, or decontaminate or replace the resistive element and tubing between subjects. Alternatively, a disposable sensor may be used. Disposable sensors, when appropriately used, avoid the need for decontamination of sensors and mouthpieces (see Disposable in-line filters section).

When an open-circuit technique (either volume or flow spirometers) is used without inspiration from the measuring system, only the mouthpiece would need to be changed or decontaminated between subjects. However, it is difficult, if not impossible, to assure that patients do not inhale through the device. A low-resistance one-way valve may be used to prevent inhalation, and, if used, must be demonstrated not to alter the spirometric measurements. Not having patients inspire through the device may make it difficult to assess test quality because of the absence of an inspiratory tracing. Hence, this technique should be used with caution. Disassembling, cleaning and/or sensor replacement will usually require recalibration of the spirometer.

Tuberculosis

In settings where TB or other diseases that are spread by droplet nuclei are likely to be encountered, proper attention to environmental engineering controls, such as ventilation, air filtration or ultraviolet decontamination of air, should be used to prevent disease transmission.

Haemoptysis and oral lesions

Special precautions should be taken when testing patients with haemoptysis, open sores on the oral mucosa or bleeding gums. Tubing and breathing valves should be decontaminated before reuse, and internal spirometer surfaces should be decontaminated with accepted disinfectants for blood-transmissible agents.

Other known transmissible infectious diseases

Extra precautions should be taken for patients with known transmissible infectious diseases. Possible precautions include the following: 1) reserving equipment for the sole purpose of testing infected patients; 2) testing such patients at the end of the day to allow time for spirometer disassembly and disinfection; and 3) testing patients in their own rooms with adequate ventilation and appropriate protection for the technician.

Disposable in-line filters

These may be an effective and less expensive method of preventing equipment contamination. The influence of commercially available in-line filters on forced expiratory measures, such as forced vital capacity (FVC) and forced expiratory volume in one second (FEV1) has not been well characterised. A low-impedance barrier device was found not to have a significant effect on FVC and FEV1 14, whereas a barrier filter has been shown to cause small but significant reductions in FEV1 (-44 mL) and peak expiratory flow (PEF; -0.47 L·s−1), but did not appear to affect DL,CO, alveolar volume or TLC 15. Although significant differences between measurements with and without filters have been demonstrated for FVC, FEV1, airway resistance and specific airway conductance (sGaw) 16, these differences were unrelated to the average values of the measurements (except for sGaw), and the limits of agreement were within the range of intra-individual short-term repeatability for almost all of the function indices. Thus, the effect of a filter with optimal characteristics is not considered to be clinically significant, and no appreciable classification error was found in diagnostic tests.

If in-line filters are used, the measuring system should meet the minimum recommendations for accuracy, precision (reproducibility), flow resistance and back pressure with the filter installed. Airflow resistance must be measured with in-line filters in place if that is how patients are tested. Manufacturers of in-line filters should provide evidence that their filter does not alter standard lung function measurements (vital capacity, FVC, FEV1, PEF, mean forced expiratory flow between 25% and 75% of FVC, TLC and DL,CO).

In the absence of evidence for infection transmission during pulmonary function testing, and the absence of a clear-cut benefit, the regular use of in-line filters is not mandated when the precautions described in the previous Prevention sections are followed.

Use of such filters is an area of controversy. On the one hand, some spirometric equipment, particularly those incorporated in multipurpose testing systems, employ valve manifolds, which are situated proximal to breathing tubes. These valve arrangements provide internal surfaces on which the deposition of expired aerosol nuclei is likely. Given their complexity, they may be difficult to disassemble and disinfect between subjects. To the extent that in-line filters have been shown to remove microorganisms from the expiratory air stream and, thus, prevent their deposition as aerosol nuclei on spirometer surfaces, their use may be indicated. On the other hand, in-line filters have been relatively inefficient in excluding microorganisms at the high flows often seen in pulmonary testing, and instrument contamination has been observed when filters have been used 17–20. However, barrier filters with a high efficiency (>99%) for excluding bacteria have been reported 21, 22, but their performance in excluding smaller microorganisms such as viruses is unknown. A reduction in overall costs with in-line filters, as compared with a disinfection approach to hygiene, in a pulmonary laboratory has been reported 17.

The use of in-line filters does not eliminate the need for regular cleaning and decontamination of lung function equipment.

Equipment design

Manufacturers of lung function equipment are encouraged to focus on designs that can be easily disassembled for cleaning and disinfection. Purchasers of pulmonary function equipment are encouraged to inquire about cleaning and disinfection issues prior to purchase of an instrument, which should involve an evaluation of the ease of cleaning and the clarity of written instructions, and an understanding of what equipment and chemicals will be required.

Level of infection risk

Lung function equipment has not been directly implicated in the transmission of infections, although there is indirect evidence of infection transmission during pulmonary function testing. Organisms from the respiratory tract of test subjects have been recovered from mouthpieces and the proximal surfaces of tubing through which subjects breathe 19, 23. The flows generated during spirometric manoeuvres may be high enough to aerosolise contaminant organisms, although such aerosolisation has not been demonstrated. There is one case report of a TB skin-test conversion following exposure to a spirometer previously used to test a patient with documented TB 24. Likewise, there is circumstantial evidence that contaminated lung function equipment may be implicated in increasing the prevalence of Burkholderia cepacia infections among cystic fibrosis patients at one centre 25. There is evidence that pneumotachometer-based systems are less susceptible to bacterial contamination than water-sealed spirometers 26. In addition, it is well documented that community water supplies can be contaminated with Mycobacteria spp. and Pseudomonas aeruginosa organisms 27–29. Thus, there is a potential for both patients/subjects and healthcare workers to deposit microorganisms onto spirometer surfaces (including mouthpieces, nose clips, tubing and any internal or external machine surface), which could subsequently come into direct or indirect contact with other patients or healthcare workers.

This does not pose an appreciable threat to patients/subjects/workers with competent immune systems. It has been argued that immunocompromised patients may require only a relatively small infective dose of either opportunistic organisms or common pathogens for infection to occur. However, there is no direct evidence that routine pulmonary function testing poses an increased risk of infection to immunocompromised patients.

Concerns for the protection of immunocompromised patients, along with increased public and provider awareness of hospital infection-control issues since the 1990s, has led many laboratory directors to routinely use in-line filters to reassure patients and laboratory personnel that their protection has been considered.

PERSONNEL QUALIFICATIONS AND TECHNICIAN'S ROLE IN QUALITY CONTROL

Personnel qualifications

Previously, the ATS has published recommendations for laboratory personnel conducting pulmonary function tests 30. Minimum requirements include sufficient education and training to assure that the technician understands the fundamentals of the tests, the common signs of pulmonary diseases and the management of the acquired pulmonary function data. The ATS also recommended that medical directors should have appropriate training and be responsible for all pulmonary function testing 31. Since these initial recommendations, pulmonary function testing equipment and procedures have become considerably more complex. The use of computers has reduced the need for routine manual measurement; however, new and more complex training issues have evolved. Many providers of pulmonary function training programmes have expanded the scope and length of training to accommodate these new needs.

The current guidelines suggest that completion of secondary education and at least 2 yrs of college education would be required to understand and fulfil the complete range of tasks undertaken by a pulmonary function technician.

For pulmonary function testing, an emphasis on health-related sciences (nursing, medical assistant, respiratory therapy, etc.) is desirable. Formal classroom-style training alone does not, however, establish competency in pulmonary function testing. Technicians who conduct pulmonary function testing need to be familiar with the theory and practical aspects of all commonly applied techniques, measurements, calibrations, hygiene, quality control and other aspects of testing, as well as having a basic background knowledge in lung physiology and pathology. In the USA, the National Institute for Occupational Safety and Health (NIOSH) has developed a model programme, and reviews and approves spirometry training courses. These 2- and 3-day courses include the fundamentals of spirometry standards and hands-on training. The workshop experience provides hands-on instruction in a small group setting with an experienced instructor. Students are expected to demonstrate their ability to properly prepare and administer a spirometric test, and demonstrate competency in other areas, such as calibration, recognition of unacceptable manoeuvres, etc.

This standard recommends training similar to the NIOSH-approved spirometry programme. Competency is demonstrated by passing a written and practical examination in the presence of an experienced instructor (i.e. hands-on testing and calibration). In Europe, training is being carried out differently in various countries. The ERS, through a specific Assembly (Assembly 9 for Allied Respiratory Professionals), regularly delivers relevant postgraduate course training at their annual Congress.

Spirometry refresher training is also recommended. Refresher training helps to ensure that testing technicians are informed of changes in spirometry standards and learn new skills. It also provides a mechanism for technicians to obtain answers to questions not foreseen during initial training. The need for refresher training has been recognised by several organisations, including the Lung Health Study 32, the National Health and Nutrition Examination Survey 33, 34 and the American College of Occupational and Environmental Medicine 35. The frequency of refresher training is dependent on many factors that differ among individuals. A recommended frequency of every 3–5 yrs is recommended, or shortly after changes to lung function standards are published. While in-house training may achieve the desired goals, laboratory directors should strongly consider the benefits of formal training programmes from outside providers.

Technician's role in quality control

Quality control is important to ensure that the laboratory is consistently meeting appropriate standards. In any quality-control programme, an important element is a manual of procedures that contains the following: calibration procedures, test-performance procedures, calculations, criteria, reference values source, and action to be taken when “panic” values are observed. A notebook or an equivalent method of recording and later producing these results should be maintained, which documents the daily instrument calibration, as well as any problems encountered with the system, corrective action required, and system hardware and software upgrades. Records of anomalous events involving either patients/subjects or the technician should be documented with the results of subsequent evaluation and responses to the event. The technician should also maintain records of continuing education and the results of evaluation and feedback provided by the medical director. The ATS has produced a complete procedure manual (Pulmonary Function Laboratory Management and Procedure Manual), which is available in paper and electronic format (www.thoracic.org/education/labmanual.asp), so as to be modifiable by laboratories to meet their individual needs.

In Europe, technical information on lung function tests is contained in a series of publications in the European Respiratory Journal 36–42.

Perhaps the most important component in successful pulmonary function testing is a well-motivated, enthusiastic technician. The importance of a quality-control programme with feedback to technicians in obtaining adequate spirometry results has been documented 32. A quality-control programme that continuously monitors technician performance is critical to the collection of high-quality data. Feedback to the technicians concerning their performance should be provided on a routine basis, which should include, at a minimum: 1) information concerning the nature and extent of unacceptable manoeuvres and nonreproducible tests; 2) corrective action that the technician can take to improve the quality and number of acceptable manoeuvres; 3) positive feedback to technicians for good performance; and 4) comments regarding system set-up and reporting results.

Manufacturers are encouraged to include quality-control aids in their software packages. However, technicians should be trained not to rely exclusively on these quality-control prompts, since technical errors may occur that are not among those recognised by the software. An example of a quality-control aid is a calibration-logging program, which stores the date, time, technician name and the results of routine daily calibration checks. Additionally, the program could issue a warning if an acceptable daily calibration check has not been performed.

REFERENCE VALUES

Detailed statements on the selection of reference values and interpretation of lung function tests have been published 39, 41–43 and new recommendations have just been created 44. In selecting appropriate reference values, it is important to choose a source that used similar equipment and had a test population that included the age range, sex and ethnic group of individuals to be tested. Also, all spirometric indices should use the same source for reference values (i.e. FVC and FEV1 should not be taken from a different reference value source than the FEV1/FVC %).

INTERPRETATION STRATEGIES

For a full account of interpretive strategies, the ATS and ERS have now revised 44 their previous statements 39, 41–43.

The interpretation of lung function tests involves two tasks: 1) the classification of the derived values with respect to a reference population and assessment of the reliability of the data; and 2) the integration of the obtained values into the diagnosis, therapy and prognosis for an individual patient.

The first task is ordinarily the responsibility of the laboratory director or his/her designee, and not only serves to communicate information to referring healthcare providers, but is also an important aspect of laboratory quality control. The second task is usually the responsibility of the physician requesting the studies and is performed within the context of patient care.

It is the responsibility of the laboratory director to develop explicit procedures for the interpretation of lung function tests and to select appropriate reference values. The procedures for interpretation and choosing reference values may legitimately vary from laboratory to laboratory, depending upon geographical location and the characteristics of the population being tested. The interpretative strategy should be consistent and take into consideration the consequences of false-positive and false-negative errors. In this way, referring physicians will not infer a change in the condition of the patient from a change in interpretation when it is, in fact, the result of a change in the approach of the interpreting physician.

ABBREVIATIONS

Table 4 contains a list of abbreviations and their meanings, which will be used in this series of Task Force reports.

Which of the following instructions should a medical assistant provide to a patient prior to spirometry testing?

  • Download figure
  • Open in new tab
  • Download powerpoint

Table. 1—

Conditions where suboptimal lung function results are likely

Table. 2—

Activities that should preferably be avoided prior to lung function testing

Table. 3—

Possible order for undertaking lung function tests in a laboratory

Table. 4—

List of abbreviations and meanings

Acknowledgments

M.R. Miller: University Hospital Birmingham NHS Trust, Birmingham, UK; R. Crapo and R. Jensen: LDS Hospital, Salt Lake City, UT, USA; J. Hankinson: Hankinson Consulting, Inc., Valdosta, GA, USA; V. Brusasco: Università degli Studi di Genova, Genova, Italy; F. Burgos: Hospital Clinic Villarroel, Barcelona, Spain; R. Casaburi: Harbor UCLA Medical Center, Torrance, CA, USA; A. Coates: Hospital for Sick Children, Toronto, ON, Canada; P. Enright: 4460 E Ina Rd, Tucson, AZ, USA; C.P.M. van der Grinten: University Hospital of Maastricht, Maastricht, the Netherlands; P. Gustafsson: Queen Silvias Children's Hospital, Goteborg, Sweden; D.C. Johnson: Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA; N. Maclntyre: Duke University Medical Center, Durham, NC, USA; R. McKay: Occupational Medicine, Cincinnati, OH, USA; D. Navajas: Lab Biofisica I Bioenginyeria, Barcelona, Spain; O.F. Pedersen: University of Aarhus, Aarhus, Denmark; R. Pellegrino: Azienda Ospedaliera S. Croce e Carle, Cuneo, Italy; G. Viegi: CNR Institute of Clinical Physiology, Pisa, Italy; and J. Wanger: Pharmaceutical Research Associates, Inc., Lenexa, KS, USA.

Footnotes

  • For editorial comments see page 1.

  • Received March 23, 2005.
  • Accepted April 5, 2005.
  • © ERS Journals Ltd

References

  1. Pulmonary terms and symbols: a report of the ACCP-ATS Joint Committee on Pulmonary Nomenclature. Chest 1975;67:583–593.

  2. International vocabulary of basic and general terms in metrology. PD 6461 Vocabulary of Metrology. Part 1: Basic and general terms (international). 2nd Edn. Geneva, International Standards Organisation, 1993

  3. American Thoracic Society. Standardization of spirometry. Am Rev Respir Dis 1979;119:831–838.

  4. Townsend MC. Spirometric forced expiratory volumes measured in the standing versus the sitting posture. Am Rev Respir Dis 1984;130:123–124.

  5. Parker JM, Dillard TA, Phillips YY. Arm span-height relationships in patients referred for spirometry. Am J Respir Crit Care Med 1996;154:533–536.

  6. Aggarwal AN, Gupta D, Jindal SK. Interpreting spirometric data: impact of substitution of arm span for standing height in adults from North India. Chest 1999;115:557–562.

  7. Chumlea WC, Guo SS, Wholihan K, Cockram D, Kuczmarski RJ, Johnson CL. Stature prediction equations for elderly non-Hispanic white, non-Hispanic black, and Mexican American persons developed from the NHANES III data. J Am Dietetic Assoc 1998;98:137–142.

  8. Chumlea WC, Guo SS, Steinbaugh ML. Prediction of stature from knee height for black and white adults and children with application to mobility-impaired or handicapped persons. J Am Dietetic Assoc 1994;94:1385–1388.

  9. Newton MF, O'Donnell DE, Forkert L. Response of lung volumes to inhaled salbutamol in a large population of patients with severe hyperinflation. Chest 2002;121:1042–1050.

  10. Akesson U, Dahlstrom JA, Wollmer P. Changes in transfer factor of the lung in response to bronchodilatation. Clin Physiol 2000;20:14–18.

  11. Botman MJ, de Krieger RA. Contamination of small volume medication nebulizers and its association with oropharyngeal colonization. J Hosp Infect 1987;10:204–208.

  12. Dautzenberg B. Prevention of nosocomial infection during nebulization and spirometry. Rev Pneumol Clin 2001;57:91–98.

  13. Tabalan OC, Williams WW, Martone WJ. Infection control in pulmonary function laboratories. Infect Control 1985;6:442–444.

  14. Denison DM, Cramer DS, Hanson PJV. Lung function testing and AIDS. Respir Med 1989;83:133–138.

  15. Johns DP, Ingram C, Booth H, Williams TJ, Walters EH. Effect of a microaerosol barrier filter on the measurement of lung function. Chest 1995;107:1045–1048.

  16. Fuso L, Accardo D, Bevignani G, Ferrante E, Della Corte A, Pistelli R. Effects of a filter at the mouth on pulmonary function tests. Eur Respir J 1995;8:314–317.

  17. Side EA, Harrington G, Thien F, Walters EH, Johns DP. A cost-analysis of two approaches to infection control in a lung function laboratory. Aust N Z J Med 1999;29:9–14.

  18. Pierce RJ. Infection control in the respiratory laboratory: risk, costs, expediency. Aust N Z J Med 1999;29:3–4.

  19. Leeming JP, Kendrick AH, Pryce-Roberts D, Smith D, Smith EC. Use of filters for the control of cross-infection during pulmonary function testing. J Hosp Infect 1993;23:245–246.

  20. Leeming JP, Pryce-Roberts DM, Kendrick AH, Smith EC. The efficacy of filters used in respiratory function apparatus. J Hosp Infect 1995;31:205–210.

  21. Kendrick AH, Milkins C, Smith EC, Leeming J, Benbough JE. Assessment of spiroguard and vitalograph bacterial filters for use with lung function equipment. Am J Respir Crit Care Med 1998;157:A58

  22. Kirk YL, Kenday K, Ashworth HA, Hunter PR. Laboratory evaluation of a filter for the control of cross-infection during pulmonary function testing. J Hosp Infect 1992;20:193–198.

  23. Rutala DR, Rutala WA, Weber DJ, Thomann CA. Infection risks associated with spirometry. Infect Control Hosp Epidemiol 1991;12:89–92.

  24. Hazaleus RE, Cole J, Berdischewsk M. Tuberculin skin test conversion from exposure to contaminated pulmonary function testing apparatus. Respir Care 1981;26:53–55.

  25. Isles A, Maclusky I, Corey M, et al. Pseudomonas cepacia in cystic fibrosis: an emerging problem. J Pediatr 1984;104:206–210.

  26. Burgos F, Torres A, Gonzalez J, Puig de la Bellacasa J, Rodriguez-Roisin R, Roca J. Bacterial colonization as a potential source of nosocomial respiratory infections in two types of spirometer. Eur Respir J 1996;9:2612–2617.

  27. Du Moulin GC, Stottmeier KD, Pelletier PA, Tsang AY, Hedley-White J. Concentration of Mycobacterium avium by hospital hot water systems. JAMA 1988;260:1599–1601.

  28. Von Reyn CF, Waddell RD, Eaton T, et al. Isolation of Mycobacterium avium complex from water in the United States, Finland, Zaire, and Kenya. J Clin Microbiolol 1993;31:3227–3230.

  29. Eichorn JH, Bancroff ML, Laasberg H, du Moulin GC, Saubermann AJ. Contamination of medical gas and water pipelines in a new hospital building. Anesthesiology 1977;46:286–289.

  30. Gardner RM, Clausen JL, Epler G, Hankinson JL, Permutt S, Plummer AL. Pulmonary function laboratory personnel qualifications. Am Rev Respir Dis 1986;134:623–624.

  31. American Thoracic Society. ATS respiratory care committee position paper: director of pulmonary function laboratory. ATS News 1978;4:6

  32. Enright PL, Johnson LR, Connett JE, Voelker H, Buist AS. Spirometry in the Lung Health Study: 1. Methods and quality control. Am Rev Respir Dis 1991;143:1215–1223.

  33. Hankinson JL, Bang KM. Acceptability and reproducibility criteria of the American Thoracic Society as observed in a sample of general population. Am Rev Respir Dis 1991;143:516–521.

  34. Hankinson JL, Odencrantz JR, Fedan KB. Spirometry reference values from a sample of the general U.S. population. Am J Respir Crit Care Med 1999;159:179–187.

  35. Townsend MC. ACOEM position statement. Spirometry in the occupational setting. American College of Occupational and Environmental Medicine. J Occup Environ Med 2000;42:228–245.

  36. Clausen JL, Coates AL, Quanjer PH. Measurement of lung volumes in humans: review and recommendations from an ATS/ERS workshop. Eur Respir J 1997;10:1205–1206.

  37. Quanjer PH, Lebowitz MD, Gregg I, Miller MR, Pedersen OF. Peak expiratory flow: conclusions and recommendations of a Working Party of the European Respiratory Society. Eur Respir J 1997;10: Suppl. 24 S2–S8.

  38. Quanjer PH, Sly PD, Stocks J. Respiratory function measurements in infants: symbols, abbreviations and units. Eur Respir J 1995;8:1039–1056.

  39. Stocks J, Quanjer PH. Reference values for residual volume, functional residual capacity and total lung capacity. ATS Workshop on Lung Volume Measurements. Official Statement of The European Respiratory Society. Eur Respir J 1995;8:492–506.

  40. Quanjer PH, Tammeling GJ, Cotes JE, et al. Symbols, abbreviations and units. Working Part Standardization of Lung Function Tests, European Community for Steel and Coal. Eur Respir J 1993;6: Suppl. 16 S85–S100.

  41. Quanjer PH, Tammeling GJ, Cotes JE, Pedersen OF, Peslin R, Yernault JC. Lung volumes and forced ventilatory flows. Report Working Party Standardization of Lung Function Tests, European Community for Steel and Coal. Official Statement of the European Respiratory Society. Eur Respir J 1993;6: Suppl. 16 S5–S40.

  42. Cotes JE, Chinn DJ, Quanjer PH, Roca J, Yernault JC. Standardization of the measurement of transfer factor (diffusing capacity). Report Working Party Standardization of Lung Function Tests, European Community for Steel and Coal. Official Statement of the European Respiratory Society. Eur Respir J 1993;6: Suppl. 16 S41–S52.

  43. American Thoracic Society. Lung function testing: selection of reference values and interpretative strategies. Am Rev Respir Dis 1991;144:1202–1218.

  44. Pellegrino R, Viegi G, Enright P, et al. Interpretative strategies for lung function tests. Eur Respir J 2005;(In press)

Which of the following instructions should a medical assistant provide to a patient prior to spirometry testing quizlet?

Which of the following instructions should a medical assistant provide to a patient prior to spirometry testing? - Smoking causes constriction an can interfere with the accuracy of spirometry testing. The patient should not smoke during the hour prior to the test.

When completing a fecal occult blood test a medical assistant is testing for?

Colorectal cancer is one of the most common types of cancer in the United States. A fecal occult blood test can screen for colorectal cancer to help find the disease early when treatment may be most effective.

Which test measures a patient's hearing using air and bone conduction?

A Rinne test evaluates hearing loss by comparing air conduction to bone conduction. Air conduction hearing occurs through air near the ear, and it involves the ear canal and eardrum. Bone conduction hearing occurs through vibrations picked up by the ear's specialized nervous system.

At which point during a patients venipuncture process should a medical assistant label the blood specimen tube?

At which point during a venipuncture process should a medical assistant label the blood specimen tube? The medical assistant should label the blood specimen tubes after the venipuncture procedure is completed, but prior to the patient being discharged.